TECHNICAL INFORMATION
The radionuclides are deposited on a 0.01 in (0.3 mm) thick aluminum substrate (anodized surface), fixed on a stainless steel support of thickness 0.1 in (2.6 mm) ensuring the rigidity of the assembly.
The radiological characteristics (emergent flux) of the sources are measured with a calibrated absolute proportional 2π sr counter.
It should be noted that the size of a calibration source should be adapted to suit the size of the detector to be controlled or calibrated : it is recommended not to use wide area sources to control detectors of a significantly different size from the detector itself. LEA does not provide any guarantee on results for other uses than those recommended.
STANDARD GEOMETRIES
Catalog references |
On request |
|
Active diameter |
||
α&β |
15 mm |
From 15 to 160 mm
|
36 mm |
||
44 mm |
||
110 mm |
||
External diameter |
||
α&β |
30 mm |
From 20 to 170 mm
|
47 mm |
||
50 mm |
||
120 mm |
||
Activity |
||
α |
0.4 kBq |
|
β |
4 kBq |
From 0.5 to 8 kBq
|
Radionuclide |
||
α |
238Pu, 239Pu, 241Am |
233U, 235U, 238U |
β |
14C, 60Co, 90Sr, 137Cs, 147Pm |
32P, 35S, 45Ca, 63Ni, 89Sr, 99Tc, 129I |
Radionuclide |
2π sr alpha flux 2π sr beta flux |
k=2 measurement uncertainty |
Equivalent activity |
Reference |
||
Under COFRAC* accreditation |
||||||
α |
241Am |
200 |
α.s-1 |
≤ 6 % |
0.4 kBq |
AM241 ESA [Type] 20 |
238Pu |
200 |
α.s-1 |
≤ 6 % |
0.4 kBq |
PU238 ESA [Type] 20 |
|
239Pu(1) |
200 |
α.s-1 |
≤ 6 % |
0.4 kBq |
PU239 ESA [Type] 20 |
|
β |
14C |
1 500 |
β.s-1 |
≤ 6 % |
4 kBq |
C14 ESA [Type] 20 |
60Co |
1 900 |
β.s-1 |
≤ 6 % |
4 kBq |
CO60 ESA [Type] 20 |
|
137Cs |
2 400 |
β.s-1 |
≤ 6 % |
4 kBq |
CS137 ESA [Type] 20 |
|
147Pm |
1 900 |
β.s-1 |
≤ 6 % |
4 kBq |
PM147 ESA [Type] 20 |
|
90Sr + 90Y |
2 500 |
β.s-1 |
≤ 6 % |
4 kBq |
SR90 ESA [Type] 20 |
|
Standard manufacturing tolerance: ± 30% IAEA Category : 5 • ISO2919 Classification: C11111 |
How to compose the reference ?Replace [Type] with the letter K, L, M or N according to the required geometry. For example: SR90 ESA K 20. |
|||||
1) Source subject to export control on dual-use goods in accordance with CE Regulation 428/2009 from council of 5 May 2009: an end user certificate will notably be requested from the customer. |
||||||
* Scope N°2-6386 available on www.cofrac.fr or upon request. |