Features
-
Scintillator-based detectors with a silicon photomultiplier
-
Ability to store data in non-volatile memory
-
Sound, light, and vibration alarms of threshold levels exceeding
-
Infrared port for transferring the dose accumulation history and event history from the dosimeter to the personal computer, as well as for setting up the dosimeter
-
Glass-filled plastic shock-resistant housing of high ingress protection rating IP67
-
Backlight of the digital display
-
Resistant to 60 repetitive shocks, each corresponding to a drop of 10 cm on a hard steel surface, and resistant to 6 drops (one drop on each side) from a height of 1 m on a hard surface
Specifications
Measurement and display range of photon ionizing radiation DER | 1∙10-6 … 10 Sv/h |
Main relative permissible error limit of EPD-27 DoseG dosimeter when measuring gamma radiation DER at 137Cs calibration with a confidence probability of 0.95
|
20 % 15 % |
Main relative permissible error limit of EPD-27 DoseGX dosimeter when measuring gamma radiation DER at 137Cs calibration with a confidence probability of 0.95
|
20 % 15 % 15 % |
Measurement and display range of X-ray radiation DER by EPD-27 DoseGX dosimeter | 1∙10-6 … 1∙10–1 Sv/h |
Measurement and display range of photon ionizing radiation DE | 1∙10-7 … 10 Sv |
Main relative permissible error limit of DE measurement at 137Cs calibration with 0.95 confidence probability, not more | 15 % |
Complementary relative permissible error limit of photon ionizing radiation DER and DE measurement result caused by ambient temperature deviation from 20 oС, in the temperature range from minus 20 to + 50 oС | 5 % per each 10 оС of deviation from 20 оС |
Operating supply voltage of the dosimeter from Li-polymer battery with a capacity of at least 400 mAh | 3.7 V |
Time of continuous operation of the dosimeter under normal climatic conditions when powered from a fully charged battery:
|
170 h
4 h
|
Unstable readings of the dosimeter during 8 hours of continuous operation, not more | 5 % |
Energy range of detected photon ionizing radiation | 0.05 … 10 MeV |
Energy dependence when measuring photon ionizing radiation DER and DE relative to 0.662 MeV energy (137Cs), not more
|
20 % 40 % |
Energy range of detected X-ray radiation by EPD-27 DoseGX dosimeter | 12 … 200 keV |
Energy dependence of EPD-27 DoseGX when measuring X-ray radiation DER and DE relative to 0.662 MeV energy (137Cs), not more | from minus 30 % to 35 % |
Operating temperature range | – 20 … + 50 °С |
Dimensions of the dosimeter with a clip, not more | 84.5 ´ 55 ´ 24.5 mm |
Weight of the dosimeter, not more | 0.11 kg |