TECHNICAL INFORMATION

Radionuclides are deposited between two hot-sealed polyester foils (approximately 75μm thick each). Each face of the source is gold vacuum-coated for optimum measurement efficiency. They are then mounted in a steel ring.

 

Radiological characteristics of the sources are measured with a calibrated gas flow proportional counter.

STANDARD GEOMETRIES

Source beta EBSA

Source beta EBSBSource beta EBSC

PRODUCTION RANGE

Catalog references

On request

Active diameter

3 mm 

From 3 to 30 mm

 

30 mm 

External diameter

25 mm 

From 25 to 60 mm

 

38 mm 

50 mm 

Activity

0.08 kBq

From 0.05 to 30 kBq

 

3 kBq 

Radionuclide

22Na, 36Cl, 60Co, 90Sr/90Y, 137Cs/137mBa, 147Pm, 204Tl, 89Sr, 134Cs

32P, 35S, 45Ca, 63Ni, 99Tc, 129I

 

 

Radionuclide

4π sr beta flux

k=2 measurement uncertainty

Equivalent activity

Reference

 
 

Under COFRAC* accreditation

 

36Cl

80

β.s-1

≤ 1.5%

0.08 kBq

CL36 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq 

CL36 EBS [Type] 30

 

60Co

80

β.s-1

≤ 1.5%

0.08 kBq 

CO60 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq 

CO60 EBS [Type] 30

 

134Cs

80

β.s-1

≤ 1.5%

0.08 kBq 

CS134 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq 

CS134 EBS [Type] 30

 

137Cs

80

β.s-1

≤ 1.5%

0.08 kBq

CS137 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq

CS137 EBS [Type] 30

 

22Na

80

β.s-1

≤ 1.5%

0.08 kBq 

NA22 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq

NA22 EBS [Type] 30

 

147Pm

80

β.s-1

≤ 1.5%

0.08 kBq

PM147 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq 

PM147 EBS [Type] 30

 

89Sr

80

β.s-1

≤ 1.5%

0.08 kBq

SR89 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq 

SR89 EBS [Type] 30

 

90Sr + 90Y

80

β.s-1

≤ 1.5%

0.08 kBq 

SR90 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq 

SR90 EBS [Type] 30

 

204Tl

80

β.s-1

≤ 1.5%

0.08 kBq 

TL204 EBS [Type] 20

 

3 000

β.s-1

≤ 1.5%

3 kBq

TL204 EBS [Type] 30

 
             

Standard manufacturing tolerance: ± 30%  

IAEA Category : 5 • ISO2919 Classification: C11111

How to compose the reference ? 

Replace [Type] with the letter A, B or C according to the required geometry. For example: TL204 EBS A 30.

 

* Scope N°2-6386 available on www.cofrac.fr or upon request.

 

KITS

EBS A and EBS B sources are also available in kit form.
Radionuclides supplied are 147Pm, 134Cs, 137Cs, 90Sr+90Y, 204Tl and 22Na with an activity of 80 Bq to3 000 Bq).
Other activities of EBS sources are available on request.